Abstract
Radiographic techniques were applied to a human simulation pelvis phantom with evaluations of the resultant image analyzed by a designated radiologist. Results indicate that a lower patient dose is received when coupled with an increase in the tube voltage. Images analyzed demonstrated little noise variation between images, which indicates the ability to lower patient dose while maintaining quality images.
Recommended Citation
Stewart, Grace, "Milliamperage-seconds, kilovoltage and patient dose in computed radiography" (2010). Master's Theses. 183.
DOI: 10.58809/AGAG2029
Available at:
https://scholars.fhsu.edu/theses/183
Rights
© 2010 Grace Stewart